The Unsung Hero of Our GC-MS: Maintaining the Electron Ionization Filament

Let's be honest - when you get data back from the lab, the last thing you're thinking about is the inner workings of the instrument used to generate it. But one tiny, unheralded component plays a huge role in making sure your analytical results are accurate: the electron ionization (EI) filament.

By: Erin Tracy and Jalice Manso


What the Heck is Electron Ionization?

Before we get to the filament, let's quickly cover the basics of EI. It's the secret sauce that powers a lot of modern mass spectrometers, including the GC-MS instrument we use to analyze your samples.

The EI process is pretty straightforward - a beam of high-energy electrons blasts apart the molecules in your sample, creating a cloud of positively charged ions. These ions then get accelerated and sorted by the mass spectrometer, allowing us to identify the compounds present. It's like breaking a puzzle into pieces and reassembling them.

The Filament That Fuels the Flame

At the heart of the EI source is a thin wire called the filament. When electricity runs through it, the filament heats up and emits the electrons that do the ionization. But over time, all that heat takes a toll, causing the filament to wear down and eventually burn out. Below you can see the spiral coil that should remind you of an incandescent bulb 💡.

Spent filament on the left, new replacement on right.

As you can see, a tired, worn-out filament (on the left) looks pretty beat up compared to a fresh, new one (on the right). If it's not replaced in time, this physical deterioration can impact the instrument's performance, leading to:

  • Weaker signals and poorer detection of your target compounds
  • Higher background noise that can make your data harder to interpret
  • Potential damage to other sensitive parts of the mass spec

Keeping Our Instrument Running Strong

Our Agilent GCMS was purchased by the North Carolina General Assembly via the NC Collaboratory using opioid settlement funds. We are grateful and feel it is our duty to keep the instrument running strong.

The good news is that replacing the EI filament is a routine maintenance task that's pretty straightforward. And we can do it ourselves! We typically anticipate swapping it out every 5-8 months to keep our GC-MS running at peak performance. After the replacement, the vacuum re-pressurization in the chamber takes about half a day. So, expect us to to be down for about a day every 6 months to do this maintenance.

Move over IKEA! Filament is #7

So the next time you get data from the lab, give a silent thank-you to the unsung hero of the GC-MS - the hardworking EI filament that makes it all possible. We are on top of taking care of the equipment that powers our science.